Time resolved stimulated Raman and Transient absorption setup
back »»| Location | Contact | |
| IAP University of Bern Sidlerstrasse 5 3012 Bern | Andrea Cannizzo Tel +41 31 631 89 37 andrea.cannizzo@iap.unibe.ch | |
Setup / Configuration | Literature | |
|
| |
The image shows the sample position (in figure the sample is replaced by an alignment pin hole) and the pathways of the actinic (blue), Raman (green) and probe (red) pulse paths (yellow figures are a guide for the eye to show the beam geometry). To compensate for shot to shot laser fluctuations, a replica of the probe pulse is detected as a reference shot per shot (dashed lines). Probe and reference are then dispersed to be detected by a single shot broadband double detector. Inset shows a zoom of the sample position. | The stimulated Raman signal is generated by stimulating the Raman emission process, induced by a ps pulse (Raman pump), with a sub 100 fs broadband pulse (probe pulse), as depicted in the figure below.
When the sample is pre-excited with a pump pulse (actinic pulse) the Raman spectrum of the excited state can be detected. |



